Skip to content

HRS 500 SpectraPro HRS

The SpectraPro HRS ensures high spectral resolution, as well as an astigmatism-corrected design for multichannel fiber applications. With dual exit ports the SpectraPro HRS can be used as two array detectors or as both a monochromator and spectrograph.

Description

SpectraPro HRS Overview

The SpectraPro® HRS delivers high spectral resolution combined with an astigmatism-corrected optical design, making it ideal for multichannel fiber applications. Featuring dual exit ports, the SpectraPro HRS can be configured with two array detectors or operated simultaneously as a monochromator and a spectrograph.

ResXtreme Spectral Deconvolution Technology

ResXtreme improves spectral resolution and peak intensity of spectral lines by up to 60%, while also delivering an overall improvement in spectral uniformity of up to 60% across all CCD positions. By significantly enhancing signal-to-noise performance, ResXtreme ensures optimal spectral results and is included as standard with all SpectraPro HRS systems.

Full-Size Dual Focal Planes

The SpectraPro HRS is equipped with two exit ports, allowing the mounting and operation of two cameras. This enables flexible operation as both a monochromator and a spectrograph within a single instrument.

AccuDrive™

AccuDrive software automatically identifies the installed turret and gratings during system start-up and performs multiple optical alignment routines to ensure accurate initialization. This results in improved wavelength accuracy and repeatability, outperforming conventional scanning systems in both precision and reproducibility.

Specifications

Specifications

Focal Length 500 mm
Aperture Ratio f/6.5
Spectral Resolution (CCD) 0.07 nm
Spectral Resolution (ResXtreme) 0.05 nm or better (typ.)
Reciprocal Linear Dispersion 1.52 nm/mm
Wavelength Coverage (26.8 mm CCD Width) 41 nm
Focal Plane Size (W × H) 30 mm × 14 mm (both exit ports)
Scan Range (1200 g/mm Grating) 0 to 1500 nm
Grating Drive Step Size 0.002 nm/step
Wavelength Accuracy ±0.2 nm (up to 0.02 nm with IntelliCal)
Turret Interchangeable triple-grating CTS turrets, self-aligning upon installation
Grating Change Repeatability 0.02 nm (typ.)
Grating Size 68 mm × 68 mm (68 mm × 84 mm optional)
Number of Turrets / Gratings Up to 3 turrets, each accommodating 3 gratings
Astigmatism Toroidal mirror design reduces astigmatism for multichannel spectroscopy
Computer Interface USB 2.0
Dimensions (L × W × H) 562 mm × 301 mm × 222 mm
Weight 22.7 kg
Optical Axis Height 156 mm ± 11 mm (adjustable)
Certification CE

SpectraPro HRS Comparison

SpectraPro HRS Comparison

Specification HRS 300 HRS 500 HRS 750
Focal Length 300 mm 500 mm 750 mm
Aperture Ratio f/3.9 f/6.5 See datasheet
Spectral Resolution (CCD) 0.10 nm 0.07 nm See datasheet
Spectral Resolution (ResXtreme) 0.07 nm or better (typ.) 0.05 nm or better (typ.) See datasheet
Reciprocal Linear Dispersion 2.38 nm/mm 1.52 nm/mm See datasheet
Wavelength Coverage (26.8 mm CCD) 64 nm 41 nm See datasheet
Focal Plane Size (W × H) 30 mm × 14 mm 30 mm × 14 mm 30 mm × 14 mm
Scan Range (1200 g/mm Grating) 0 to 1500 nm 0 to 1500 nm 0 to 1500 nm
Grating Drive Step Size 0.002 nm/step 0.002 nm/step 0.002 nm/step
Wavelength Accuracy ±0.2 nm (up to 0.02 nm with IntelliCal) ±0.2 nm (up to 0.02 nm with IntelliCal) ±0.2 nm (up to 0.02 nm with IntelliCal)
Turret Interchangeable triple‑grating CTS turrets, self‑aligning upon installation
Grating Change Repeatability 0.02 nm (typ.) 0.02 nm (typ.) 0.02 nm (typ.)
Grating Size 68 mm × 68 mm (68 mm × 84 mm optional)
Number of Turrets / Gratings Up to 3 turrets, each accommodating 3 gratings
Astigmatism Correction Toroidal mirror design for multichannel spectroscopy
Computer Interface USB 2.0
Dimensions (L × W × H) 399 × 279 × 222 mm 562 × 301 × 222 mm See datasheet
Weight 19.5 kg 22.7 kg See datasheet
Optical Axis Height 156 mm ± 11 mm (adjustable)
Certification CE

Documentation

Download / View Documentation

Teledyne SpectraPro Datasheet

Request Quote

Request a quote









    Back To Top